错误 ID: 6719, 6720
The actual encoder pattern is checked against a known pattern.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Encoder was replaced and has a different pattern or no reset of pattern was made. | Reset pattern via user interface and recalibrate. |
2 | Encoder lost position completely. | Exchange the monochromator assembly. |
base-id: 3604124683
id: 3604124683