错误 ID: 6713, 6714
A mechanical shock to the instrument during operation may cause a movement of the monochromator. The position is lost and the lamp will turn off.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Short mechanical shock. | re-ignite the lamp. |
2 | Message appears intermittently without mechanical shock. | Exchange the monochromator assembly. |