错误 ID: 2692
Wait for temperature or wait for defined signal has not been fulfilled within the specified time frame.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Time too short. | Increase time. |
2 | The detector is waiting after the automatic purge of the reference cell. | Allow the wait time to elapse. |
base-id: 3596601355