可能的原因 | 建议采取的措施 | |
---|---|---|
1 | The diode balance value is outside the pre-set range -0.5 to +0.5 , an unequal amount of light is falling on the two light receiving diodes. | Flush the reference cell with the mobile phase being used. Perform the RID Optical Balance procedure (see The Optical Balance Procedure ). |
base-id: 3596555147
id: 3596555147