The cuvette is used for off-line measurements (no flow system required) and is basically a standard flow cell with a few changes:
wide bore capillary connections for easier injections with a syringe
identification lever for cell auto-recognition system.
Install the cuvette instead of the standard flow cell.
Connect the waste tubing to the outlet of the cuvette.
Use the syringe (see Cuvette Kit ) to inject the compound.
Setup the parameters for the Fluorescence Scan (under Special Setpoints).
Select “Take Fluorescence Scan” on the user-interface to start the off-line measurement.