错误 ID: 6704
The lamp current of the xenon flash lamp is monitored constantly. If the current gets too high, an error is generated and the lamp is turned OFF.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Short-circuit of trigger pack assembly or defective FLL board. | 请与您的 Agilent 服务代表联系。 |
2 | Short-circuit of flash lamp assembly. | 请与您的 Agilent 服务代表联系。 |
base-id: 3586630027