The visible lamp voltage is missing.
The processor continually monitors the voltage across the lamp during operation. If the lamp voltage falls below the lower limit, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | 电源出现故障。 | 请与您的 Agilent 服务代表联系。 |
2 | Defective connector or cable. | 请与您的 Agilent 服务代表联系。 |
base-id: 3359653387
id: 3359653387