对于生物兼容性模块,只能使用生物 / 生物兼容部件! 请勿与生物惰性部件混用。 |
适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | 8710-0510 | 开口扳手 1/4-5/16 英寸 |
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | G4767-60500 | 校准过的样品定量环,100 µL,右侧(蓝色编码) (for G7167C) OR |
1 | ![]() | G7137-60500 | 样品定量环 MP35N 100 µL,右侧(蓝色/橙色编码) (for G7137B) |
前提条件
|
警告
没有罩的进样针有造成人身伤害的危险
没有罩的进样针有对操作人员造成伤害的危险。
小心
样品定量环配置不匹配
系统损坏
小心
Blockages inside of the needle assembly union
Use a 1/4-inch wrench to tighten the fitting of the loop capillary.
Close the front door.
In the Local Controller close Change loop.
OR
In the Agilent Lab Advisor software Change Loop, click Back and wait until the needle is in the needle park station.
注意
Choose the Change Needle procedure, if you need an auto referencing step included.
注意
If you have changed the sample loop, verify that the correct sample loop is configured in the CDS, see Setting Up the Autosampler With OpenLab ChemStation .
For G7167C and G7137B, proceed with the subsequent steps.
Perform Hydraulic Path Leak Test.
Perform Sampler Leak Test if needed for troubleshooting in case of Hydraulic Path Leak failure.