The following pages describe the maintenance (simple repairs) of the module that can be carried out without opening the main cover.
Procedure | Typical interval (minimum) |
---|---|
Change needle/needle seat | 30000 needle into seat movements |
Change peristaltic pump cartridge | 3000 h on time |
Change rotor seal | As needed |
Change metering seal | 30000 injections |
Procedure | Typical interval (minimum) |
---|---|
Change needle/needle seat | 60000 needle into seat movements |
Change peristaltic pump cartridge | 3000 h on time |
Change rotor seal | As needed |
Change metering seal | 30000 injections |
Procedure | Typical interval (minimum) |
---|---|
Change Needle/Needle Seat | 60000 needle into seat movements |
Change Rotor Seal | As needed |
Change Metering Seal | 30000 injections |
Change Stator or entire Valve Head | 10000 injections (approx. 30000 switches) |