错误 ID: 25409
The autosampler failed to complete initialization correctly. The autosampler initialization procedure can not move the motors inside of the sample handler to their reference positions in a predefined routine. During initialization, the processor monitors the position sensor and encoders to check for correct movement. If one or more of the movements or the status information of the subparts is not read out successfully, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Sample handler is blocked. | Check if the sample handler can move freely. Switch off the instrument. Do an autoreferencing (the needle assembly must be installed in the needle parkstation during this procedure). |
2 | Defective sample handler motors. | 请与您的 Agilent 服务代表联系。 |
base-id: 3803727371