错误 ID: 25034
The autosampler failed to complete the alignment correctly.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Mechanical obstruction of the sample handler. | Ensure unobstructed movement. |
2 | Defective sample handler motors. | 请与您的 Agilent 服务代表联系。 |
base-id: 3803574027
id: 27021601567797003