错误 ID: 30823
The heating function of the Integrated Column Compartment is switched off because the trend analysis of the reading from the digital temperature sensor has failed.
To verify the correct functioning of the heating element, perform the Heater Test in the Agilent Lab Advisor software.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Connection problems. | Power off the system and verify that the ICC is correctly mounted in the sampler. Power on the system and check if the problem persists. |
2 | Potential hardware error. | 请与您的 Agilent 服务代表联系。 |
base-id: 3716085387