错误 ID: 30820
The system reported an error because reaching the set temperature took longer than expected. To verify the correct functioning of the heating, perform the Heater Test in the Agilent Lab Advisor software.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Connection problems. | Power off the system and verify that the ICC is correctly mounted in the sampler. Power on the system and check if the problem persists. |
2 | Potential hardware error. | 请与您的 Agilent 服务代表联系。 |
base-id: 3716051083
id: 18014402225533067