错误 ID: 30717
The system is in an error state because the calibration of the analog temperature sensor has failed.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Sampler incompatibility. | If the hosting sampler is a Vialsampler, verify its compatibility with the Sample Cooler installed. Units with the serial number DEBAT02000 or below are equipped with an analog temperature sensor that is not compatible with the Vialsampler. |
2 | Potential hardware error. | 请与您的 Agilent 服务代表联系。 |
base-id: 3715842059
id: 18014402225324043