错误 ID: 34320
The analytical head is not operational because there is no valid information available for its type in the firmware.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Missing configuration data. | Verify that the type of the analytical head is correctly configured in the CDS or Lab Advisor. |
2 | Potential hardware error. | 请与您的 Agilent 服务代表联系。 |