错误 ID: 34319
The injection valve position is not found or the valve is unable to switch.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Overtightened screws. | Verify that the stator screws are not overtightened as it can lead to hindered rotor movements. If necessary, loosen the stator screws and retighten them appropriately. |
2 | Broken rotor seal. | Check the rotor seal for signs of visible damage and if necessary, replace it. |
3 | Broken stator. | Check the stator for signs of visible damage and if necessary, replace it. |
4 | Potential hardware error. | 请与您的 Agilent 服务代表联系。 |