错误 ID: 34224
The home position of the metering plunger is not found or the plunger is unable to move.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Blockage in the flow path. | Check if there is a significant pressure difference (> 10 – 15 bar) between the mainpass and the bypass position of the injection valve. If yes, systematically replace the components of the mainpass: Needle seat, Needle, Rotor seal, Sample loop |
2 | Potential hardware error. | 请与您的 Agilent 服务代表联系。 |
base-id: 3715216651
id: 18014402224698635