エラー ID:
ASTM Noise Test Evaluation
考えられる原因 | 対策 | |
|---|---|---|
1 | Insufficient lamp warm-up time. | Allow detector and UV-lamp turned on for at least 2 hours. |
2 | Absorbing solvent or air bubble in flow cell. | 確実にフローセルが水で満たされ、気泡がないようにします。 |
3 | Dirty or contaminated flow cell. | Flush flow cell Clean the flow cell as described in Clean the Max-Light Cartridge Cell. Clean the flow cell as described in Maintenance of Standard, Semi-Micro or Micro Flow Cell. |
4 | UV ランプが古い。 | UV ランプを交換します。 |
5 | Environment not according to specifications. | Improve environment. |