WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
Open the front door.
NOTE
This graphic shows the Hybrid Multisampler (G7167C).
Check the condition of the needle assembly and the sample loop. Replace them if necessary, see Exchange the Needle Assembly and Remove the Sample Loop-Flex.
NOTE
Take care that the needle is installed properly. The plastic adapter must be installed correctly and the sample loop should not be kinked.
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
Unlock the needle.
NOTE
This procedure is completely different than the standard PM replacement of the needle assembly in Lab Advisor. The safety lock of the needle assembly has to be released by carefully sliding the pusher upwards.
Verify that the needle assembly is unlocked after installation.
Reset the sampler (using the instrument control) or turn the instrument Off/On again to start the initialization.
Close the front door.
Wait until the initialization of the sampler is completed.
If the error persists, contact your local service representive.