The dimensions and weight of your module (see Specifications) allow it to be placed on almost any laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches) at the rear for the circulation of air and electric connections.
The G4301A SFC Control Module should be operated in a vertical, the other modules in a horizontal position.
If a Column Compartment is installed, an additional 25 cm (10 inches) of space on either side for the circulation of air, and approximately 8 cm (3.1 inches) at the rear is required for electrical connections.
If a complete 1260 Infinity III system is to be installed on the bench, make sure that the bench is designed to carry the weight of all the modules. For a system including the Thermostatted Autosampler it is recommended to position the modules in two stacks.