The Agilent 1260 Infinity III SFC Control Module is heavy (approximately 26 kg or 56 lbs). Enlist the aid of a co-worker to share the lifting load in order to avoid possible injury. Select the laboratory bench space before your system arrives. Pay special attention to the total height requirements. Avoid bench space with overhanging shelves. Pay special attention to the total weight of the modules and solvents you have in addition to the G4301A SFC Control Module. Make sure that your laboratory bench can support this weight.
Type | Specification | Comments |
---|---|---|
Weight | 26 kg (56 lbs) |
|
Dimensions (height × width × depth) | 60 cm x 26 cm x 48 cm (23 in x 10 in x 18 in) |
|
Line voltage | 100 - 240 VAC, ±10 % |
|
Line frequency | 50 - 60 Hz, ±5 % |
|
Power consumption | 700 VA max |
|
Operating temperature | 15 – 30 °C (59 – 86 °F) |
|
Non-operating temperature | -40 – 70 °C (-40 – 158 °F) |
|
Humidity | <95 %, at 25 – 40 °C | Non-condensing |
Laboratory ventilation | Minimum 6 air exchanges/hr for lab air; | CO₂ monitor recommended w/ alarm @5000 ppm |
Exhaust vent capacity | >20 L/min with sustained negative pressure |
|
Operating altitude | up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage Category II, Pollution Degree 2 | For Indoor Use only |
ISM Classification | ISM Group 1 Class A | According to CISPR 11 |
Type | Specifications |
---|---|
Inlet CO₂ bulk purity | >99.99 % vapor; >99.999 % liquid |
Inlet CO₂ phase | vapor from non-dip-tube high pressure cylinder; liquid from commercial CO₂ delivery system |
Inlet CO₂ supply pressure | 40 - 70 bar (580 - 1000 psi) |
Inlet CO₂ temperature | 15 - 30 °C (59 - 86 °F) |
Liquid coolant | 30 % propylene glycol in deionized water; proprietary antioxidants; red dye added for safety |
Coolant volume | < 280 mL |
Type | Specifications |
---|---|
High pressure flow path | 300 and 400 series stainless steel, carbon filled PEEK, UHMW PE, sapphire, ceramic, gold |
Low pressure flow paths [waste, wash pump, leak tray]Only when a legacy wash pump is installed | 316 stainless steel, PEEK PTFE, FEP, CTFE CPE; LDPE Tygon PVC |
Vapor exhaust | Tygon PVC |
[1]Only when a legacy wash pump is installed.
Type | Specifications |
---|---|
Hydraulic system | Single piston with proprietary motor control |
Total hydraulic volume | <5 mL at pressure <70 bar <25 mL at pressure up to 400 bar |
Chiller system | Thermoelectric cooling with secondary air/liquid cooling circuit |
Chiller temperature | -20 – 9 °C (-4 – 48.2 °F) |
Back-Pressure Regulation (BPR) system | Low volume diaphragm type with proprietary drive control; replaceable BPR head assy; No recalibration required after head replacement |
Booster pump pressure range | 100 – 400 bar up to 5 mL/min demand |
BPR thermal range | 40 – 70 °C (104 – 158 °F) |
BPR thermal precision | ±1 K |
BPR settable pressure range | 100 – 400 bar |
Back-pressure precision | ±0.25 bar |
Back-pressure thermal precision | ±1 K |
Programmable back-pressure gradient | Yes. Up to 20 timetable entries. |
Communications | USB 2.0; APG Remote: ready, start, stop and shut-down signals; relay contact closure (wash pump only) |