The Agilent 1260 Infinity III SFC Control Module requires approximately 30 cm (1 ft) of linear bench space immediately adjacent to the target 1260 Infinity III system stack. Approximately 13 cm (5 inches) of free space is required behind the instrument for cable access and adequate air flow for ventilation. Similar access to the rear of the system is also required to install cables and interface cards. For optimal performance, the rear air space should not be heated significantly above room temperature by the exhaust of other instrumentation in the lab. Hot exhaust should be vented or directed upward from the instrument.
The module is designed to be installed on either side of the system stack with sufficient high pressure transfer tubing to attach to a double-stacked system. If the system is attached to a split-flow detector such as Mass Spec or ELSD, the G4301A SFC Control Module should be positioned on the opposite side of the stack. Shelves overhanging the G4301A SFC Control Module should provide a minimum of 15 cm (6 inches) of clearance to allow access to the rear on/off switch. Finally, the PC system interface to the G4301A SFC Control Module is USB 2.0. A 1.8 m (6 ft) cable is supplied with the system. The CPU must be placed within range of this cable. Alternately, the user may supply an extended length USB cable not to exceed 5 m (16 feet).
While the G4301A SFC Control Module can exist on either side of the 1260 Infinity III system, it is often easier to locate it on the left side. For more specification details, see Specifications.