ASTM: “Evaporative Light Scattering Detectors Used in Liquid Chromatography”.
Gas flow | 1.6 SLM |
Neb temperature | 70 °C |
Evaporator temperature | 70 °C |
Data Rate | 40 Hz |
Smoothing | 1 |
ASTM drift tests require a temperature change below 2 °C/h (3.6 °F/h) over one hour period. Our published drift specification is based on these conditions. Larger ambient temperature changes will result in larger drift. Better drift performance depends on better control of the temperature fluctuations. To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 °C/h (1.8 °F/h). Turbulences around one minute or less can be ignored. ASTM measurements require that the detector should be turned on enough time before start of testing.