For bio-inert modules use bio-inert parts only! |
| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G4267-60300 | Sample Loop Flex 20 µL, right (red coded) |
1 | ![]() | G4267-60400 | Sample Loop Flex 40 µL, right (green coded) |
1 | ![]() | G4267-60500 | Sample Loop Flex 100 µL, right (blue coded) |
1 | ![]() | G7167-68500 | Sample Loop Cartridge 500 µL right |
1 | ![]() | G7167-68900 | Sample Loop Cartridge 900 µL right |
1 | ![]() | G5668-60500 | Bio-inert Sample Loop 100 µL (for G5668A) |
1 | ![]() | G7137-60300 | Sample Loop MP35N 20 µL, right (red/orange coded) |
1 | ![]() | G7137-60400 | Sample Loop MP35N 40 µL, right (green/orange coded) |
1 | ![]() | G7137-60500 | Sample Loop MP35N 100 µL, right (blue/orange coded) |
Further sample loops for the Dual-Needle option are available, see Sample Loops and Capillaries (Dual Needle).
Prerequisites
|
WARNING
Risk of injury by uncovered needle
An uncovered needle is a risk of harm to the operator.
In the Agilent Lab Advisor software select Service & Diagnostics > Maintenance Positions > Change Needle, Loop and Seat , click Start and wait until the needle assembly is in maintenance position.
OR: In the Local Controller start the maintenance mode and select Change Needle, Loop and Seat function.
The needle assembly is still connected to the loop capillary. Use a 1/4 inch wrench to loosen the fitting of the loop capillary connected to the analytical head.
Lock the needle in the safety position.
CAUTION
Damage of the loop
The loop shape may be damaged if the loop is stretched or bent too far.
WARNING
Sharp needle
Uncovered needles may cause injuries
Remove the needle assembly by slightly pulling the needle cartridge.
Remove the cartridge out of its proper position. By gently tilting and pulling it out of the work space of the multisampler.