Your module will work within specifications at ambient temperatures and relative humidity as described in:
ASTM drift tests require a temperature change below 2 °C/hour (3.6 °F/hour) over one hour period. Our published drift specification is based on these conditions (refer to Specifications). Larger ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature fluctuations. To realize the highest performance, minimize the frequency and the amplitude of the temperature changes to below 1 °C/hour (1.8 °F/hour). Turbulences around one minute or less can be ignored.
CAUTION
Condensation within the module
Condensation can damage the system electronics.
NOTE
This module is designed to operate in a typical electromagnetic environment, i.e. where RF transmitters such as mobile telephones may not be used in close proximity.
base-id: 3610920843
id: 9007202865661835