Parameter Name | External Value | Comment |
---|---|---|
Weight | 14.7 kg (32.4 lbs) |
|
Dimensions (height × width × depth) | 180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 80 VA, 70 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification |
---|---|
Detection type | Refractive Index |
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting |
Refractive index range | 1.00 – 1.75 RIU, calibrated |
Measurement range | ± 600·10-6 RIU |
Optical zeroing | Digitally via software, manually via set screw |
Optics temperature control | 5 °C above ambient to 55 °C |
Sample cell | Volume: 2.5 µL Maximum pressure: 5 bar (0.5 MPa) Maximum flow rate: 1 mL/min (100 % water) |
Valves | Automatic purge and automatic solvent recycle |
Volumes | Inlet port to sample cell 2.5 µL, inlet port to outlet port 265 µL |
Liquid contact materials | 316 stainless steel, PTFE and quartz glass |
pH range | 2.3 – 9.5 |
Performance specifications | Short term noise: <±1.75·10-9 RIU Drift: <200·10-9 RIU/h |
Time programmable parameters | Polarity, peak width |
Maximum data rate | 148 Hz |
Detector zero | Automatic zero before analysis |
Instrument control | LC and CE Drivers A.02.12 or above Instrument Control Framework (ICF) A.02.03 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Agilent Instant Pilot (G4208A) with firmware B.02.19 or above Lab Advisor B.02.07 or above |
Analog output | Recorder/integrator: 100 mV or 1 V, output range selectable, one output |
Communication | LAN, Controller Area Network (CAN) Extended Remote Interface (ERI): ready, start, stop and shut-down signals |
Safety and maintenance | Extensive diagnostics, error detection and display with Agilent InfinityLab Assist and with Agilent Lab Advisor software. Leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
GLP | Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-selectable limits and feedback messages. Electronic records of maintenance and errors. Automated operational qualification/performance verification (OQ/PV). |
Housing | All materials recyclable. |
base-id: 8040449291