When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Laboratory balance |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| DAB/Ph Eur/BP/JP/NF/USP Grade Sucrose |
1 | ![]() | 5190-1539 | Syringe 5mL with Luer-Lock |
1 | ![]() | 9301-0407 | Syringe, External Valve adapter, SST |
1 | ![]() | 5190-5111 | Syringe filter, 0.45 µm, 100/pk |
1 | ![]() | 0100-1516 | Finger-tight fitting PEEK, 2/pk |
Preparation of the sucrose calibration solution (final concentration: 3.5 g/L).
Wait five minutes and shake again. The solution is now ready for use.
Preparing the pump.
Flush the degasser and pump.
Run the RID Tools Screen with the Agilent Lab Advisor (for further information see Online-Help of user interface). (If an Agilent pump is part of the system, the pump section is active.)
Purging the sample and reference cells.
Fill the sample cell with calibration solution.
NOTE
Do not inject the calibration solution without the sample filter.
Leave the RID Tools Screen.