The detector described in this manual is designed for highest optical performance, GLP compliance and easy maintenance. It includes the following features:
Maximum of 120 Hz (G7117A/G7117C) or 240 Hz (G7117B) data acquisition rate
Higher sensitivity for conventional LC as well as ultra fast applications by using next generation optical design
Increased sensitivity with 60 mm Max-Light cartridge flow cell
Optimized cell geometry for less peak dispersion for narrow bore applications
Max-Light cartridge flow cells for standard applications are available, see Max-Light Cartridge Flow Cell
More reliable and robust peak integration process (automated) due to less baseline noise/drift/refractive index and thermal effects especially under ultra fast gradient conditions
RFID tracking technology is used for the UV-lamp and the Max-Light cartridge flow cells
Multiple wavelength and full spectral detection at 120 Hz (G7117A/G7117C)/240 Hz (G7117B) sampling rate, keeping up with the analysis speed of ultra-fast LC
Programmable 1 – 8 nm slit (G7117B) or fixed 4 nm slit (G7117A/C) for rapid optimization of sensitivity, linearity and spectral resolution provides optimum incident light conditions
Improved Electronic temperature control (ETC) provides maximum baseline stability and practical sensitivity under fluctuating ambient temperature and humidity conditions
Additional diagnostic signals for temperature and lamp voltage monitoring
Easy exchange of flow cell by cartridge design