The detector allows easy access to flow cells via a cartridge. A variety of optional flow cells can be inserted using the same quick, simple mounting system.
Max-Light Cartridge Flow Cells for standard and bio-inert applications are available. For testing of the detector, a Max-Light Cartridge Test Cell is available.
Parts list
p/n | Description | |
---|---|---|
![]() | G4212-60008 | Max-Light Cartridge Cell (10 mm, V(σ) 1.0 μL) |
![]() | G4212-60007 | Max-Light Cartridge Cell (60 mm, V(σ) 4.0 μL) |
![]() | G4212-60032 | HDR Max-Light Cartridge Cell (3.7 mm, V(σ) 0.9 μL) |
![]() | G4212-60038 | ULD Max-Light Cartridge Cell (10 mm, V(σ) 0.6 μL) |
![]() | G7117-60020 | Max-Light Cartridge Cell LSS (10 mm, V(σ) 1.0 μL) |
![]() | G4212-60011 | Max-Light Cartridge Test Cell |
The optical principle of the Max-Light Cartridge cell is based on opto-fluidic waveguides. Nearly 100 % light transmission is achieved by utilizing total internal reflection in a non-coated silica fiber. Compromising refractive index and thermal effects are almost completely eliminated, resulting in significantly less baseline drift.
NOTE
For additional information on the Max-Light Cartridge flow cell refer to Choosing a Flow Cell and Inline Pressure Relief Valve Kit (G4212-68001).