Several flavors of the Max-Light Cartridge Flow Cell are available, see Specifications for Max-Light Cartridge Flow Cells.
Cartridge Cells |
|
Maximum pressure | 70 bar (1015 psi) Maximum Operating Pressure (MOP)Maximum operating pressure (MOP)2 150 bar (2175 psi) Maximum Incidential Pressure (MIP)Maximum incidental pressure (MIP)2 |
pH range | 1.0-12.5 (solvent dependent) |
[1]Maximum operating pressure (MOP): Maximum pressure at which a system can operate continuously under normal conditions.
[2]Maximum incidental pressure (MIP): The maximum pressure which the system can experience during a short time.
If higher sensitivity is necessary, the G4212-60007 (Max-Light Cartridge Cell (60 mm, V(σ) 4.0 μL)) can be used. This cell enhances the detector by lowering the limit of detection (LOD) by a factor of about 3 (depending on the application).
The G4212-60008 (Max-Light Cartridge Cell (10 mm, V(σ) 1.0 μL)) covers a wide range of applications:
all column diameter down to at least 2.1 mm ID or even less
applications with peak dispersion (Peakwidth x flow) down to ~2 µL [example: pw = 0.04 min at flow = 0.1 mL/min gives peak dispersion of 0.04 min x 0.1 mL/min = 0.004 mL = 4 µL]
The Max-Light Cartridge ULD cell can be used with the G7117A DAD FS and G7117B DAD. The cell is a requirement for the Ultra-Low Dispersion Kit solution which currently exists as 5067-5189 (InfinityLab Ultra-Low Dispersion Kit) . The cell should be part of the ultra-low dispersion solution.
The Max-Light Cartridge HDR cell can be used with the G7117A DAD FS and G7117B DAD. The cell is required as a part of the High Dynamic Range (HDR) solution.
NOTE
To protect the flow cell against overpressure (e.g. in systems with LC/MS) install G4212-68001 (Inline Pressure Relief Valve Kit) , see Inline Pressure Relief Valve Kit (G4212-68001).