Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Flow cell |
For details on flow cells see:
NOTE
The shown cell parts will differ depending upon the flow cell type. For detailed parts schematics, refer to above mentioned pages.
1 | Cell screw | 2 | Conical springs |
3 | Ring #1 PEEK | 4 | Window Quartz |
5 | Gasket #1 (small hole) | 6 | RFID tag |
7 | Gasket #2 (large hole) | 8 | Ring #2 PEEK |
Disassembling the Flow Cell.
CAUTION
Scratched window surfaces by tweezers
Window surfaces can easily be scratched by using tweezers for removing the windows.
Cleaning the Flow Cell Parts
NOTE
Always use new gaskets.
Reassembling the Flow Cell
NOTE
The semi-micro #1 and #2 gaskets (items 6 and 7, Semi-micro Flow Cell (Parts)) look very similar. Do not mix them up.
Repeat the procedure for the other cell side.
Reconnect the capillaries.
Perform a leak test. If OK, insert the flow cell.
{Comment by , : ToDo: Condition: "not(($ProductGroup='G7114B'))" }Perform --- MISSING LINK --- to check the correct positioning of the flow cell.
Perform Wavelength Calibration to check the correct positioning of the flow cell.
Replace the front cover.
base-id: 3284790923
id: 9007202539531915