The detector described in this manual is designed for highest optical performance, GLP compliance and easy maintenance. It includes the following features:
Higher data rate up to 120 Hz (G7114A) or 240 Hz for ultra-fast-HPLC (G7114B)
Deuterium lamp for highest intensity and lowest detection limit over a wavelength range of 190 to 600 nm
Optional flow-cell cartridges (standard 10 mm, 14 µL; high pressure 10 mm, 14 µL; micro 3 mm, 2 µL; semi-micro 6 mm, 5 µL), Bio, and Prep Cells are available and can be used depending on the application needs (other types may be introduced later)
Dual wavelength mode, see Dual-Wavelength Mode
Easy front access to lamp and flow cell for fast replacement
Electronic identification of flow cell and lamp with RFID (Radio Frequency Identification) tag for unambiguous identification
Lamp information: part number, serial number, production date, ignitions, burn time
Cell information: part number, serial number, production date, nominal path length, volume, maximum pressure
Built-in electronic temperature control (ETC) for improved baseline stability
Built-in holmium oxide filter for fast wavelength accuracy verification
base-id: 3283444107
id: 9007202538185099