The detector provides a Dual-Wavelength mode that offers additional operation functions.
Features:
200 ms acquisition time per data point
5 Hz data rate distributed to two channels
2.5 Hz data rate for each channel,
delta wavelength max. 150 nm,
scans during Dual-Wavelength mode are possible,
the second order filter is disabled when one wavelength is < 370 nm.
Timetable:
Wavelength settings are timetable programmable (depends if enough time for implementation is available),
switching from Single-Wavelength mode to Dual-Wavelength mode is NOT timetable programmable,
filter settings are not timetable programmable.
base-id: 3283501195