Advantages:
High sample throughput
Increased productivity
High efficiency
|
|
Gradient elution is frequently used for fast separation of complex samples in LC. Since the gradient elution requires the column to regenerate before subsequent runs, an automated column regeneration system saves valuable analysis time. Agilent's InfinityLab Quick Change 2-position/10-port valve enables the simultaneous analysis of one sample on one LC column while an extra regeneration pump flushes and equilibrates a second, identical column. At the end of the run, the valve switches to the second position and the next sample is separated on the previously flushed and equilibrated column. Meanwhile, the regeneration pump flushes and equilibrates the first column. Up to 50 % of analysis time is often required to equilibrate columns. Using alternating column regeneration saves time and provides higher sample throughput.