The interconnection of ports at particular valve position strongly depends on the combination of valve and module. The software user interface always displays the correct situation. A method modification or re-plumbing of the connections is typically required if transferring methods from G1316A/B/C to G7116A/B, G1170A or G4227A.
Refer to the table below for further information on which ports are connected to which position.
Modules | Valve | Position 1 | Position 2 |
---|---|---|---|
G1316A/B/C | 2-position/6-port | 1-2 | 1-6 |
G7116A/B, G1170A, G4227A | 2-position/6-port | 1-6 | 1-2 |
G1316A/B/C | 2-position/10-port | 1-2 | 1-10 |
G7116A/B, G1170A, G4227A | 2-position/10-port | 1-10 | 1-2 |