When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | 5182-3466 | Torx screwdriver T10 (for the Sample Thermostat) OR |
1 | ![]() | 5023-3089 | Torx key set (part of the G7120-68708 InfinityLab LC Series Tool Kit) |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G7167-60201 | Sample Thermostat |
Prerequisites
|
WARNING
Flammable refrigerant
Formation of flammable gas-air mixtures inside the Sample Thermostat and laboratory.
WARNING
Flammable refrigerant used
WARNING
In the event of a damage
WARNING
Heavy weight
The module is heavy.
CAUTION
Routing of the condensation tubing
Proper routing of the condensation tubing is critical for correct condensate drainage.
CAUTION
Condensate inside the module
Damage to the electronics of the module
Ensure that no condensate remains inside the thermostat before proceeding forward.
NOTE
Gently tapping on the sides of the sampler can help to remove the last traces of condensate from the system.
Loosen the screw (1) and remove the condensate funnel (2) from the back of the thermostat.
NOTE
If there is still some condensate inside the thermostat, place a suitable container underneath the outlet tube, and keep tapping on the sides of the sampler until no more water comes out.
Pull the thermostat halfway out, disconnect the power and the data cable and then remove the unit completely from the sampler.
Damage to the cables. Do not bend or pinch the cables. Fit in the Sample Cooler/Sample Thermostat perfectly.
CAUTION
Damage to the cables.
Slide the cooler/thermostat all the way into the sampler, making sure that the cables don’t get jammed between the metal parts.
Position the condensate collector funnel underneath the condensate drainage outlet tube (1) and fix it to the back of the thermostat by tightening the screw (2). Ensure correct orientation and avoid overtightening the screw.
NOTE
For information on proper condensate handling, see G5668A, G7137A, G7167A, G7167B Multisampler Install the Sample Thermostat.
Position the condensate collector funnel underneath the condensate drainage outlet tube (1) and fix it to the back of the thermostat by tightening the screw (2). Ensure correct orientation and avoid overtightening the screw.
NOTE
For information on proper condensate handling, see Install the Sample Thermostat for the G7129A Vialsampler, G7129B Vialsampler, G7129C Vialsampler.
CAUTION
Damage to the Sample Thermostat
Switch on the sampler and perform the Sample Cooler Function Test to verify the correct functioning of the new thermostat, see Sample Cooler Function Test.