The Agilent InfinityLab Sample Thermostat is the combination of an electric heater and a vapor-compression refrigeration system. It uses isobutane as a non-Freon refrigerant, which is harmless to the environment and does not affect the ozone layer and global warming, but it is combustible. Please adhere to the warnings listed in the manual.
Type | Specification | Comments |
---|---|---|
Weight | < 6 kg (< 13.2 lbs) |
|
Dimensions (height x width x depth) | 205 x 340 x 370 mm (8.1 x 13.4 x 14.6 inches) |
|
Refrigerant gas | R600a (max. 0.030 kg) | Ozone depletion potential (ODP) =0 Global warming potential (GWP) =3 |
Supply voltage | 24 VDC |
|
Current | 10 A max. |
|
Ambient operating temperature | 4 °C to 40 °C (39 °F to 104 °F) | For sample cooling, ambient temperature ≥10 °C |
Ambient non-operating temperature | -40 °C to +70 °C (-40 °F to +158 °F) |
|
Humidity | < 95 % r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
CAUTION
General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
Type | Specifications |
---|---|
Operating principle | High performance, low-energy consumption micro-compressor based cooler with natural R600a coolant (Isobutane max. 0.030 kg), user-upgradable |
Temperature range | from 4 °C to 40 °C |
Temperature settable | from 4 °C to 40 °C in 1 ° increments |
Temperature accuracy (<25 °C, <50 % r.H.) | 2 °C to 6 °C at a setpoint of 4 °C |
Type | Specification |
---|---|
LC & CE Drivers | A.02.14, (A.02.18)Minimum version for full thermostat functionality. or above |
Instrument Control Framework (ICF) | A.02.04, (A.02.05)Minimum version for full thermostat functionality. or above |
Lab Advisor Software | 2.19 or above |
Firmware | D.07.37 or above |
[1]Minimum version for full thermostat functionality.
base-id: 4796537483
id: 45036001070242443-3