Type | Specification | Comments |
---|---|---|
Weight | 14.5 kg (32 lbs) |
|
Dimensions (height × width × depth) | 180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 80 VA, 65 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Hydraulic system | Dual piston in series pump with servo-controlled variable stroke drive, power transmission by gears and ball screws, floating pistons |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting | |
Flow range | Settable: 0.001 – 10 mL/min Recommended: 0.2 – 10.0 mL/min | Set points in 0.001 mL/min increments For use with flow rates below 500 μL/min a vacuum degasser is required. |
Flow precision | ≤ 0.07 % RSD, or <0.02 min SD, whichever is greater | Based on retention time at constant temperature |
Flow accuracy | ± 1 % or 10 µL/min, whichever is greater | Pumping degassed H₂O at 10 MPa (100 bar, 1450 psi) |
Pressure operating range | Up to 60 MPa (600 bar, 8702 psi) up to 5 mL/min Up to 20 MPa (200 bar, 2901 psi) up to 10 mL/min |
|
Pressure pulsation | < 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar, 44 psi), whichever is greater | At 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 145 psi) |
Compressibility compensation | User-selectable, based on mobile phase compressibility |
|
Recommended pH range | 1.0 – 12.5 | Solvents with pH < 2.3 should not contain acids which attack stainless steel |
Integrated degassing unit | Optional Number of channels: 2 Internal volume per channel: 1.5 mL |
|
Instrument Control | LC & CE Drivers A.02.14 or above Instrument Control Framework (ICF) A.02.04 or above Lab Advisor B.02.08 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Instant Pilot (G4208A) with firmware B.02.20 or above | For details about supported software versions refer to the compatibility matrix of the LC & CE Drivers version in use |
Communication | Controller Area Network (CAN) Enhanced Remote Interface (ERI) Local Area Network (LAN) Universal Serial Bus (USB) |
|
Safety and maintenance | Extensive diagnostics, error detection and display through Agilent Lab Advisor, leak detection, safe leak handling, leak output signal for shutdown of the pumping system. Low voltage in major maintenance areas. |
|
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors |
|
Housing | All materials are recyclable |
NOTE
For use with flow rates below 500 μL/min a vacuum degasser is required.
base-id: 4290031243