Probable cause | Suggested actions | |
---|---|---|
1 | Check the number of injections does not exceed the typical number of injections for this column and application. Replace the column if necessary. | |
2 | Verify that the column chemistry is suitable for the given analytical problem. Switch to another column type if necessary. | |
3 | Ensure that the method settings are properly set and the gradient system is delivering the correct gradient composition. Check if the mobile phase reservoirs are empty or if the filter frits are blocked. Check for air bubbles in the system. If necessary, purge the pump and system with a viscous mobile phase to remove any air bubbles. Check the system for any leaks. Check for leaking pump seals. If necessary, replace pump seals. Check for sticking check valves. If necessary, sonicate or replace check valves. | |
4 | Flush and, if necessary and allowed by the supplier, backflush the column. Replace the column if necessary. If using a guard column or inline filter, replace it. | |
5 | Column overload | Reduce the injection volume or dilute the sample. Switch to a column with a higher sample capacity (increase length or diameter). |
6 | Improper method settings | Check temperature settings in the column compartment. |