Probable cause | Suggested actions | |
---|---|---|
1 | Determine the column void volume and system dwell volume to determine the optimal equilibration time for the gradient method. | |
2 | Check the number of injections does not exceed the typical number of injections for this column and application. Replace the column if necessary. | |
3 | Ensure that the method settings are properly set and the gradient system is delivering the correct gradient composition. Check if the mobile phase reservoirs are empty or if the filter frits are blocked. Check for air bubbles in the system. If necessary, purge the pump and system with a viscous mobile phase to remove any air bubbles. Check the system for any leaks. Check for leaking pump seals. If necessary, replace pump seals. Check for sticking check valves. If necessary, sonicate or replace check valves. | |
4 | Switch to a fresh batch of solvents. | |
5 | Column/mobile phase incompatibility with the analyte | Change the chemistry of the column to strengthen interactions with the analyte. Choose the pH of the eluent that is not too close to the pKa value of the analyte. Ensure that the pH of the eluent within the column manufacturer’s specifications. |
6 | Insufficient flushing after use of buffer | Extensively flush the system with water prior to switching from methods using buffer to methods using organic solvents. |
base-id: 4302968203
id: 4302968203