Probable cause | Suggested actions | |
---|---|---|
1 | Mobile phase absorption/refraction too high (UV Detectors) | Use less absorptive/refractive solvents or mobile phase additives. Change UV wavelength or switch the detector polarity . |
2 | Injection solvent effects | If possible, use the starting mobile phase composition or weaker solvents as the injection solvent. Ensure that the injection solvent is miscible with the mobile phase and causes no salt precipitation. |
3 | Improper method settings | Check the detector method settings , such as wavelength and data rate, and optimize them if needed. |