Probable cause | Suggested actions | |
---|---|---|
1 | Extra-column effects | Check your system for long tubing lengths between components and where possible, replace them with shorter connection lengths or capillaries with smaller internal diameters. |
2 | Ensure that your fittings are made correctly | |
3 | Column overload | Reduce the injection volume or dilute the sample. Switch to a column with a higher sample capacity (increase length or diameter). |
4 | Injection solvent effects | If possible, use the starting mobile phase composition or weaker solvents as the injection solvent. Ensure that the injection solvent is miscible with the mobile phase and causes no salt precipitation. Reduce the injection volume. |
5 | Check the number of injections does not exceed the typical number of injections for this column and application. Replace the column if necessary. | |
6 | Large dwell volume | If possible, reduce the dwell volume of your system. Check if the affected peak is eluted during the gradient, and not isocratically. If necessary, reduce the initial gradient composition to focus peaks. |
7 | Improper method settings | Check the detector method settings , such as wavelength and data rate, and optimize them if needed. |
8 | Worn injection valve rotor seal | Replace the rotor seal in the injection valve. |
9 | Ensure that the choice of the mobile phase is appropriate for the analyte. Ensure solvent lines are associated with the desired mobile phases. If in doubt switch to a fresh batch of solvents. | |
10 | Wrong column | Ensure that stationary phase selection is appropriate for the analyte. Ensure mobile phase and column compatibility. Ensure using columns with wider pores (≥ 300 Å) for larger molecules and smaller pores (< 300 Å) for small molecules. |
base-id: 4302704779
id: 4302704779