Probable cause | Suggested actions | |
---|---|---|
1 | Column overload | Reduce the injection volume or dilute the sample. Switch to a column with a higher sample capacity (increase length or diameter). |
2 | Reverse the column direction if allowed by the supplier. If all peaks show poor peak shape, replace the column. | |
3 | Injection solvent effects | If possible, use the starting mobile phase composition or weaker solvents as the injection solvent. Ensure that the injection solvent is miscible with the mobile phase and causes no salt precipitation. |
4 | Column/mobile phase incompatibility with the analyte | Choose the pH of the eluent that is not too close to the pKa value of the analyte. Ensure that the pH of the eluent within the column manufacturer’s specifications. |