Probable cause | Suggested actions | |
---|---|---|
1 | Ensure that the method settings are properly set and the gradient system is delivering the correct gradient composition. Check if the mobile phase reservoirs are empty or if the filter frits are blocked. Check for air bubbles in the system. If necessary, purge the pump and system with a viscous mobile phase to remove any air bubbles. Check the system for any leaks. Check for leaking pump seals. If necessary, replace pump seals. Check for sticking check valves. If necessary, sonicate or replace check valves. | |
2 | Ensure mobile phase and column compatibility. Ensure solvent lines are associated with the desired mobile phases. Use gradient elution that ends at a relatively high organic composition, such as 95 % acetonitrile, to ensure that all components are eluted during the analysis. Increase the starting organic composition, use a higher flow rate, or increase the overall strength of the eluent system to get the components to elute quicker from the column. | |
3 | Aging detector lamp | Check if the on-time hours of the detector lamp have exceeded the expected lifetime. Run detector lamp intensity test to check the condition of the lamp. Replace the lamp if necessary. |
4 | Improper method settings | Check the detector method settings , such as wavelength and data rate, and optimize them if needed. Check temperature settings in the column compartment. |
5 | Small detector cell | Make sure the detector cell has the same path length as the one used for the expected chromatogram. |
6 | Strong retention | Try a less hydrophobic stationary phase for the column. Use columns with PEEK hardware and implement the use of InfinityLab Deactivator Additive or similar when working with metal-binding analytes. |
7 | Ensure that the vials are not filled to the top. |
base-id: 4301893003
id: 4301893003