Probable cause | Suggested actions | |
---|---|---|
1 | Ensure that the system is properly purged with the mobile phase. Degas mobile phases. Purge the pump and system with a viscous mobile phase to remove any air bubbles. | |
2 | Sticking check valve | Observe the pressure plot and if you notice consistent pressure fluctuation, sonicate the check valve with a suitable solvent, such as isopropanol, or replace it. |
3 | Ensure that the method settings are properly set and the gradient system is delivering the correct gradient composition. Check if the mobile phase reservoirs are empty or if the filter frits are blocked. Check for air bubbles in the system. If necessary, purge the pump and system with a viscous mobile phase to remove any air bubbles. Check the system for any leaks. Check for leaking pump seals. If necessary, replace pump seals. Check for sticking check valves. If necessary, sonicate or replace check valves. | |
4 | Leaking seals | Perform a pump leak rate test or equivalent to determine if seals are leaking. If so, replace pump seals. |