Probable cause | Suggested actions | |
---|---|---|
1 | Plugged filter frit | Inspect filter frits in the pumps to ensure they are not clogged, replace if necessary. Inspect inline filter frits to ensure they are not clogged, replace if necessary. Inspect guard column frit to ensure it is not clogged, replace if necessary. |
2 | Flush and, if necessary and allowed by the supplier, backflush the column. Replace the column if necessary. | |
3 | Systematically loosen the fittings of the system working upstream from the waste to the pump and observe the pressure changes. Use the pressure changes while loosening the fittings to determine which tube is clogged. Replace tubing if necessary. | |
4 | Column diameter/particle size too small |
(https://www.agilent.com/en/product/liquid-chromatography/hplc-advisor) Ensure that mobile phase viscosity is suitable for the column particle size. Reduce the column length or increase the particle size. |
5 | Mobile phase viscosity too high | Use a lower viscosity solvent or reduce the flow rate. Use a higher column temperature. Reduce the column length or increase the particle size. |
6 | Ensure mobile phase compatibility with buffer. Flush the system with water to re-dissolve any salt crystals formed in the system. | |
7 | Column misuse | Determine if the mobile phases are compatible with the column manufacturer's solvent compatibility information. When changing mobile phases, ensure solvent miscibility. |