Type | Specification | Comments |
---|---|---|
Weight | 21.0 kg (46.3 lbs) |
|
Dimensions (height × width × depth) | 200 x 396 x 436 mm (7.9 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 210 VA / 180 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Hydraulic system | Two dual pistons in series pump with proprietary servo-controlled variable stroke design and smooth motion control for active damping. |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting | |
Pump resolution step size | 300 pL |
|
Flow range | Settable: 0.001 – 5 mL/min, in 0.001 mL/min increments (executed in 300 pL/step increments) |
|
Flow precision | ≤ 0.07 % RSD or 0.005 min SD, whichever is greater | Based on retention time at constant temperature |
Flow accuracy | ± 1 % or ± 10 µL/min, whichever is greater |
|
Pressure operating range | Up to 130 MPa (1300 bar) at 0 – 2 mL/min ramping down to 80 MPa (800 bar) at 5 mL/min |
|
Pressure pulsation | < 1 % amplitude or < 0.5 MPa (5 bar), whichever is greater |
|
Compressibility compensation | Automatic |
|
Recommended pH range | 1.0 – 12.5 | Solvents with pH < 2.3 should not contain acids which attack stainless steel |
Gradient formation | High-pressure binary mixing |
|
Delay volume | As low as 45 µL (10 µL without mixer) |
|
Composition range | 0 – 100 % |
|
Composition precision | < 0.15 % RSD or 0.01 min SD, whichever is greater |
|
Composition accuracy | ± 0.35 % absolute |
|
Number of solvents | 2 out of maximum 26 solvents |
|
Solvent selection valve | Internal 4-solvent selection valve included. External 2 x 12 solvent valve as option, fully integrated in the pump control interface. |
|
Integrated degassing unit | Number of channels: 2 Internal volume per channel: 1.5 mL |
|
Materials in contact with solvent | TFE/PDD copolymer, PFA, PEEK, FEP, ETFE, Stainless Steel, FFKM, Gold, Ruby, Zirconium Oxide-based ceramic, Tantalum, UHMW-PE, Polyimide, DLC Coated Stainless Steel |
|
Automatic purge valve | Included |
|
Active seal wash | Included |
|
Intelligent System Emulation Technology (ISET) | Included |
|
Instrument control | LC & CE Drivers A.02.11 or above Instrument Control Framework (ICF) A.02.03 or above Lab Advisor software B.02.06 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above Instant Pilot (G4208A) with firmware B.02.17 or above | For details about supported software versions refer to the compatibility matrix of your version of the LC & CE Drivers |
Communication | Controller Area Network (CAN), RS232C, APG remote: ready, start, stop, and shutdown signals, LAN |
|
Safety features and maintenance | Leak detection, safe leak handling, leak output signal for shutdown of the pumping system. Low voltage in major maintenance areas. Extensive diagnostics, error detection and display with Agilent InfinityLab Assist and with Agilent Lab Advisor software. |
|
GLP feature | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal wear and volume of pumped mobile phase with pre-defined and user settable limits and feedback messages. Electronic records of maintenance and errors. |
|
Housing | All materials are recyclable |
|
base-id: 5032081675