The figure shows the default settings.
Lamps: can be turned
ON/OFF.
Analog Output Range: can be set to either
100 mV or 1 Vfull scale, for additional settings see
Analog Output (under Method Parameter
Settings).
UV Lamp Tag
Automatic
detects a lamp with RFID tag. If no RFID tag lamp is used, “UV lamp not ready”
is displayed and it cannot be ignited. A compatible mode has to be selected
based on the used lamp; see Non-RFID-tag lamp information below.
Manual (by PN) uses the selected “heating”
mode. This mode can also be used when the RFID tag of the standard lamp
(5190-0917) is not recognized (defect RFID tag or
reader).
Non-RFID-tag lamp: In case a non-RFID-tag
lamp is used, the user interface will show this when selecting a compatible
mode. You may operate the detector outside of the guaranteed specification. The
correct selection is important for optimal performance and
lifetime.
At Power On: automatic lamp-on at power
on.
Automatic Turn On: automatic detector power
on.
|