Slit Test Evaluation
Probable cause | Suggested actions | |
---|---|---|
1 | Air bubble in flow cell. | Flush the flow cell or remove the flow cell. |
2 | Old lamp. | Run the “Intensity Test”. Exchange the lamp if old or defective. |
3 | Defective slit assembly. | Please contact your Agilent service representative. |
4 | Defective detector main board. | Please contact your Agilent service representative. |
5 | Defective PDA/optical unit. | Please contact your Agilent service representative. |