When
|
WARNING
Heavy weight
The module is heavy.
CAUTION
Condensate inside the module
Damage to the electronics of the module
CAUTION
Unsecured transportation
Mechanical damage
NOTE
Moving the sampler with the Sample Thermostat installed is possible for short distances (for example, from one workbench to another). For longer transportation, remove the thermostat from the sampler and handle the units separately.
Remove condensate.
NOTE
Do not tilt the module to avoid damage to the internal parts.
Remove all sample containers from the sample hotel.
Move the robot arm to the park position using Instant Pilot or Lab Advisor, see Arm Position.
Turn off the sampler.
Install all parts of the Transport Protection, see Install the Transport Protection Foam.
Remove the Sample Thermostat from the sampler if needed, see Replace the Sample Thermostat.